The FDA has identified this as a Class I recall, the most serious type of recall. Use of this device may cause serious injuries or death.
Recalled Product:
Device Use
Photo of MitraClip Clip Delivery System®
The Abbott Vascular MitraClip Delivery System is intended to treat patients with degenerative mitral regurgitation (DMR) a condition involving a dysfunction of the heart's mitral valve. The MitraClip Clip Delivery System is indicated for use in patients who have been determined to be at prohibitive risk for mitral valve surgery.
The delivery system has three parts: a delivery catheter; a steerable sleeve; and an implantable clip. The implantable clip is introduced into the left atrium of the heart through the steerable sleeve and the delivery catheter. The implantable clip is then positioned and closed between the leaflets that separate the left atrium and the left ventricle to reduce the reversed blood flow.
Reason for Recall
Abbott Vascular has received reports of cases where the Clip Delivery System could not be detached from the Clip due to a malfunction of the device. These cases resulted in open heart surgery to retrieve the delivery system. Abbott Vascular is therefore recalling the MitraClip Delivery System to provide updated instructions and training for health care providers who use the device.
The use of affected products may cause serious adverse health consequences, including serious patient injury or death. Currently there are 3,534 devices on the market, with nine reports of this malfunction. There has been 1 death.
Who May be Affected
What to Do
On February 4, 2016, Abbott Vascular issued a safety notice to all physicians using the device instructing them to:
Contact Information:
Customers can contact Abbott Vascular Customer Support at 1-800-227-9902.
Date Recall Initiated:
February 4, 2016