  • Título: Chimney and Periscope Grafts to Facilitate Endovascular Treatment of Aortic Transection in a Patient With Aberrant Right Subclavian Artery
  • Fecha: 12-03-2014
  • Sandra Vicente, MD1,2; Michael Glenck, MD3; Dieter Mayer, MD1; Frank J. Veith, MD4;

    Mario Lachat, MD1; and Felice Pecoraro, MD5

    1Clinic for Cardiovascular Surgery and 3Interventional Radiology, University Hospital Zurich, Switzerland. 2Department of Vascular Surgery, Virgen de la Salud Hospital, Toledo, Spain. 4New York University Medical Center, New York, New York, and The Cleveland Clinic, Cleveland, Ohio, USA. 5Vascular Surgery Unit, University of Palermo, AOUP ‘‘P. Giaccone,’’ Palermo, Italy.

    Purpose: To report the use of parallel grafts to extend the proximal landing zone for stent- graft repair of aortic transection involving an aberrant right subclavian artery (ARSA).

    Case Report: A 28-year-old patient was referred for treatment of traumatic aortic transection with contained rupture at the level of an ARSA. Immediate thoracic endovascular aortic repair (TEVAR) was planned because of hemodynamic instability. To achieve rapid sealing and maintain perfusion to both subclavian arteries, a chimney stent to the left subclavian artery (LSA) and a periscope stent-graft to the ARSA were deployed successfully. After surgical repair of all fractures, the patient was discharged 1 month after the initial injury in good condition. Imaging follow-up at 10 months showed a stable repair, patent parallel grafts, and no complications.

    Conclusion: TEVAR with chimney and periscope grafts proved to be a safe and quick treatment for a patient requiring ARSA repair in acute aortic transection. This technique maintained blood flow to the ARSA and LSA in a totally endovascular approach, which could be very valuable in transection cases where bypass surgery to supra-aortic branches is compromised or deemed challenging due to thoracic wall and/or neck trauma. Parallel grafting can be a valuable tool to address any acute aortic pathology as it can be performed with off-the-shelf devices.

    J Endovasc Ther. 2014;21:123–126

    Key words: thoracic aorta, aortic transection, trauma, thoracic endovascular aortic repair, aberrant right subclavian artery, left subclavian artery, stent-graft, parallel graft, chimney graft, periscope graft, landing zone.

    Endovascular repair of traumatic aortic tran- section is an attractive alternative to open repair,1,2 but there is little experience in patients presenting aberrant right subclavian artery (ARSA).3 In some of these patients, aortic sealing might require landing a stent-

    graft proximal to the left subclavian artery (LSA). In such cases, revascularization of one or both subclavian arteries might be consid- ered. We report an endovascular approach to maintain blood flow to the ARSA and LSA that could be very valuable in transection cases where bypass surgery is compromised or deemed challenging due to thoracic wall and/or neck trauma.


    Figure: Preoperative coronal (A) and sagittal (B) CT scans show the aortic transection (arrow) localized distal to an ARSA origin (asterisk). (C) Intraoperative control angiography showing good anatomical and functional results (asterisk shows ARSA repair; arrow shows LSA stent). (D) Postoperative CT angiography showed complete sealing of the rupture site and unrestricted chimney stent (arrow) and periscope graft (asterisk shows ARSA repair). (E) CT angiography reconstruction (posterior view) performed at 10 months shows patent LSA chimney and ARSA periscope endografts.

    CASE REPORT

    A 28-year-old patient was transferred to our tertiary university center for treatment of polytrauma, including multiple bone frac- tures, thoracic trauma, and traumatic aortic transection in the classic location. Redo computed tomographic (CT) angiography af- ter transport showed an aortic lesion that the subclavian arteries were accessed with soft guidewires (J wire; Cordis Endovascular, a Johnson & Johnson company, Miami Lakes, FL, USA) via 8-F and a 9-F sheaths (Arrow International Inc., Reading, PA, USA). The ARSA was cannulated transfemorally using a JB1 catheter (Cook Medical, Bloomington, IN, USA). A 9 x100-mm Viabahn stent-graft (W.L. Gore & Associates Inc., Flagstaff, AZ, USA) was placed in the ARSA, and an 8340-mm Wallstent (Boston Scientific, Natick, MA, USA) was introduced over a Rosen wire (Cook Medical) and temporarily parked in the LSA. A 283100-mm CTAG (W.L. Gore & Associates) was introduced through a 22-F Arrow sheath and positioned under fluoroscopy and angio- graphic control to land just distal to the left common carotid artery. After the aortic stent- graft was successfully deployed at the intend- ed position, the chimney and periscope grafts were positioned precisely and deployed. Full modeling of all stents and stent-grafts was performed with a kissing balloon technique. Nonetheless, both the chimney (Wallstent) and periscope (Viabahn) devices showed significant residual constriction from the aor- tic stent-graft, which was relieved by addi- tional stenting (8337-mm Visi-Pro; Covidien, Mansfield, MA, USA). Completion angiogra- phy and manometry showed good anatomical and functional results (Figure, C). A low-flow type II endoleak was observed but not treated as the hemodynamics were stable at this stage. The postoperative course was uncom- plicated and pre-discharge CT angiography showed complete sealing of the rupture site, with patent chimney and periscope grafts (Figure, D). At 10-month follow-up (Figure, E), the patient was asymptomatic and the CT showed no complications.

    DISCUSSION

    ARSA, originating as the most distal aortic arch vessel, is the most common congenital anomaly of the aortic arch, with an incidence of 0.5% to 1.5%.5 ARSA pathologies have traditionally been treated by conventional open repair, which carries considerable mor- tality (11% to 24%) even in the largest published series.6 Recently, hybrid repair has been reported.5–8 In these hybrid procedures, seemed to be stable (Figure, A,B), but the patient developed progressive hemodynamic instability. As a consequence, it was decided to treat the transection as an emergency with a minimally invasive technique.

    Bilateral percutaneous transfemoral remote accesses were obtained using the preclose technique with Proglide (Abbott Vascular, Redwood City, CA, USA), while remote access to the LSA was achieved with infraclavicular exposure of the axillary artery and a surgi- close technique.4 After systemic heparin was administered (5000-unit intravenous bolus), endovascular vessel occluders combined with carotid-to-subclavian artery bypass grafting/ or fenestrated and aortic stent-graft deploy- ment were successful.7–11 Our case illustrates the applicability of thoracic endovascular aortic repair (TEVAR) with chimney and periscope grafts to address traumatic aortic transection in a patient presenting with ARSA. Unlike other cases in which a periscope graft has been used to extend the distal landing zone,12 we used the periscope to extend the proximal seal zone.13 This maneuver allowed us to use only one chimney endograft in this area instead of two, which reduced the potential for endoleak and long-term compli- cations. As a technical note, if a pressure drop along the chimney and/or periscope endo- grafts or severe stenosis (~80%) is detected during the procedure, additional branch stent- ing (self-expanding or balloon-expandable) should be performed.

    CONCLUSION:

    TEVAR with the chimney and periscope grafts proved to be a safe and quick treatment in a patient requiring ARSA repair in acute aortic transection. This technique maintained blood flow to the ARSA and LSA in a totally endovascular treatment. Parallel grafting can be a valuable tool to address any acute aortic pathology as it can be performed with off-the- shelf devices.

    Acknowledgment: The authors thank Dr. med. Thomas Pfammatter (Institute of Diagnostic and Interventional Radiology, University Hospital of Zurich, Switzerland) for his continuous clinical support in treating TEVAR cases.

    REFERENCES:

    1. Bruninx G, Wery D, Dubois E, et al. Emergency endovascular treatment of an acute traumatic rupture of the thoracic aorta complicated by a distal low-flow syndrome. Cardiovasc Intervent Radiol. 1999;22:515–518.

    2. Lachat M, Pfammatter T, Witzke H, et al. Acute traumatic aortic rupture: early stent-graft re- pair. Eur J Cardiothorac Surg. 2002;21:959–963.

    3. Reynolds TS, Donayre CE, Somma CG, et al. Endovascular management of blunt aortic injury with an associated aberrant right subcla-vian artery: a report of three cases. Ann Vasc Surg. 2011;25:979, e977-912.

    4. Mayer D, Rancic Z, Wilhelm M, et al. Improved hybrid technique for vascular access and closure. J Endovasc Ther. 2008;15:322–325.

    5. Morris ME, Benjamin M, Gardner GP, et al. The use of the Amplatzer plug to treat dysphagia lusoria caused by an aberrant right subclavian artery. Ann Vasc Surg. 2010;24:416, e415-418

    6. Kieffer E, Bahnini A, Koskas F. Aberrant subclavian artery: surgical treatment in thirty- three adult patients. J Vasc Surg, 1994;19:100– 110.

    7. Kedora J, Grimsley B, Pearl G. Endovascular treatment of an aberrant right subclavian artery aneurysm with use of the Zenith iliac plug. Proc (Bayl Univ Med Cent). 2009;22:144–145.

    8. Shennib H, Diethrich EB. Novel approaches for the treatment of the aberrant right subclavian artery and its aneurysms. J Vasc Surg. 2008;47: 1066–1070.

    9. Yang C, Shu C, Li M, et al. Aberrant subclavian artery pathologies and Kommerell’s diverticu-lum: a review and analysis of published endovascular/hybrid treatment options. J En- dovasc Ther. 2012;19:373–382.

    10. Ktenidis K, Lioupis A, Giannopoulos A, et al. Management of traumatic aortic isthmus rup- ture in case of aberrant right subclavian artery (arteria lusoria). Ann Vasc Surg. 2012;26:421 e421-425.

    11. Gafoor S, Stelter W, Bertog S, et al. Fully percutaneous treatment of an aberrant right subclavian artery and thoracic aortic aneu- rysm. Vasc Med. 2013;18:139–144.

    12. Rancic Z, Pfammatter T, Lachat M, et al. Periscope graft to extend distal landing zone in ruptured thoracoabdominal aneurysms with short distal necks. J Vasc Surg. 2010;51:1293– 1296.

    13. Pecoraro F, Pfammatter T, Mayer D, et al. Multiple periscope and chimney grafts to treat ruptured thoracoabdominal and pararenal aor- tic aneurysms. J Endovasc Ther. 2011;18:642– 649.




  • Fuente: endovascular.es